1. CB6 (MEDLINE
®
CARELINE
®
)
The CB6 control box has been specially developed for use together with LA27 actuator in the care
and rehab industry.
The control box is designed to be mounted on the actuator LA27 as with the CB9 and LA31 system.
The CB6 control box has a LED power ON indicator, detachable mains cable and strain relief for all
cables.
The control box CB6 communicates with the LA27 actuator by means of the built-in end stop signal
switches in the actuator. Due to the signal switches the power to the motor will be cut off in the
control box and not in the actuator.
This ensures that it is only possible to run in the opposite direction once the actuator has reached the
end position.
Usage:
Duty cycle: 2/18; 2 minutes continuous use followed by 18 minutes not in use
CB6 is approved according to EN60601-1 / UL60601-1
The CB6 can only be combined with LA27 and HB30 / HB70 / HL70 / HB80 and HL80.
Recommendations:
• Be aware of the handset configuration (e.g. CH1, 2, 3 should be CH1, 3, 4).
• Same cables variants as for CB9 CARELINE and CB12.
Connecting the system:
• Do not connect the mains cable until all actuators have been connected to the control box.
• Start by connecting the handset to the control box. The connection at the control box is marked with “HB”.
• Connect the different actuators to the different channels on the control box. Each channel is marked with a number (e.g. “1”, “2”, “3”....).
• Check that all plugs are well connected and firm pushed into the connection plug.
Due to the fact that LINAK control boxes are designed for a high IP degree, a firm force can be required.
• CONNECT the mains and turn on the power!
• Finally, connect the battery (BA18) with special T-cable or normal battery cable depending on the specified system.
• The actuators can now be operated by pushing a button on the handset. Use only one button at a time.
If the control box is equipped with special software, an initialising process might be necessary. This process is described in the software specification.
Attention should be paid to the following:
• Control boxes must only be connected to the mains voltage specified on the label.
• The control box must be connected in such a way that the cables are not trapped, exposed to tension or sharp objects, when the application is
moved in different directions.
CB6 system diagram