DK Digital CDB-700 Manual do Utilizador Página 91

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 90
Page 91 of 108
Safety concept:
DC power is only on the application when needed – when the hand control is activated.
The user is part of a safety concept with light indications showing:
Connected to the mains
Green light when connected to the mains
Power request
Yellow light at power request when the hand control is activated
Check of functionality
Yellow light is turned off when releasing a hand control button – otherwise the
hand control or other components in the application are defective. If so, the user
must call for service.
The SMPS30 is designed for placing on the floor, but it can also be mounted on the wall / application via the bottom base plate
and 4 screws.
If the SMPS30 is mounted on the wall/application, please dismount the SMPS30 to get access to the cable locking mechanism on the mains cable.
Recommendations
Please note
The SMPS30 is only compatible with LA43 IC and the LA44 IC.
The SMPS30 is not suitable for outdoor applications
Before installation, re-installation or troubleshooting
• Stop the application
• Switch off the power supply and pull out the mains plug
• Relieve the application of any loads, which may be released during the work
Before start-up
• Make sure that the system has been installed as instructed in the User Manual
• System connection. The individual parts must be connected before the SMPS30 is connected to the mains.
During operation
• Ensure that the cables are not damaged
• Unplug the SMPS30 before moving the equipment
Device protection:
The SMPS30 contains several mechanisms to protect itself from being damaged due to excessive use.
• In case of overheating, the device will activate a thermal protection. No power output will be available until the temperature is back again within
normal operating range.
• In case of exceeding the current limit / failure in the actuator, the device will activate an overload protection. Immediately after the situation has
been remediated the power output will be available again.
Hot plugging
• It is not allowed to remove or to add the output cable as long as the SMPS is powered by mains.
Maintenance/cleaning
• The SMPS30 must be cleaned at regular intervals to remove dust and dirt and it must be inspected for mechanical damage, wear and breaks
• It is not allowed to spray directly with a high-pressure cleaner on the device
• Interconnecting cables must remain plugged in, correctly fitted with O-rings, during cleaning to prevent ingress of water
• The SMPS30 is resistant to the majority of cleaners and disinfectants used in the hospital and nursing home sector.
However, the detergents must comply with the following basic requirements
- They must not be highly alkaline or acidic (pH value 6-8)
- They must not contain caustic agents