3. CBJ Care (MEDLINE
®
CARELINE
®
)
The control boxes CBJ Care is a part of the JUMBO system. JUMBO is a
modular system combining an actuator, a control box (CBJ Care/CBJ1/
CBJ2), a battery (BAJ1/BAJ2), and a charger (CHJ2)/internal charger in a
flexible solution, specially developed for patient lifts.
The complete system will be medically approved and contains a series
of features, which meet the patients need for a safe and comfortable
lift, e.g. CBJ Care is equipped with a soft-start/stop function, emergency
lowering function, learn mode for current cut off, service indicator and
battery indicators. Furthermore usage information is collected in the
control box.
Usage:
• Duty cycle: max. 10% or 2 min.continuous use then18 min. without use
• Ambient temperature: +5° to +40°
• For up to three actuators (lift, leg spreader and tilt of the sling)
Approvals: IEC60601-1:2005 3
rd
edition, ANSI / AAMI ES60601-1:2005 3
rd
edition, CAN/CSA-22.2 No 60601-1:2008
Australia deviation, for one, two or three actuators, (lift, leg spread and tilt actuator), Canadian deviation
Recommendations:
Before start-up we recommend to reset the service counters - days and cycles until next service visit. To reset press the up and
down button on the control box or the hand control for 5 seconds. An audio signal will confirm the resetting.
• The green battery indicator (100% to 50% capacity remaining) will light up during charging even though the battery is not fully charged.
It is necessary to use the CHARGE” diode to indicate whether or not the battery is fully charged (when using internal charger).
The CHARGE indicator will light up during charging and turn off when the battery is fully charged.
• When charging, the CBJ-Care will not be able to operate any actuators.
• Using the Learn Mode function means that:
The lift will never be able to lift more than 1.5 times the max. load. However the actuator will not stop exactly at the weight it is adjusted to.
This is because the actuator will use less current when it’s components are run-in. After the max. current value has been registered using the
Learn function the control box will be able to use max. current +10%.
This ensures that the lift is capable of lifting the promised load but cannot lift more than 1.5 times this amount. If the below mentioned conditions
are fulfilled:
- Learn mode is made on the specific pair of actuator and control box which are going to be used in real life afterwards.
- Ambient temperature should be app. 20 degrees C +/ - 5%.
- Load curve: difference between the highest and lowest load should not be more than approx. max. 10%
The registration function can only be activated by using a specially produced handset (HB7x235-00). A standard handset cannot activate the function.
• The maximum cut-off value that can be registered (stored) is 11 Amp.
If an actuator or CBJ Care is exchanged it will be necessary to reset the max. load to ensure the correct cut-off value for the new system as a whole
• Please note that when you disconnect the service data tool from JUMBO Care it will take 1 hour before the control box will power down.
This means that display indicators will keep lighting up 1 hour after disconnection of the service data tool. If you want to avoid this battery
consumption the emergency stop button can be activated.
• Ensure that you have the Service data tool version 2.5.0 or newer as only these versions support JUMBO Care 2
nd
generation.
• If the control box is exchanged the actuator data will be lost. The data is stored in the control box, not the actuator. However, actuator data can be
maintained by using the service data tool for saving actuator data in the new control box.
• Note that the control box has to be “awake” in order to connect to the service data tool. Activate the control box before connecting via the
emergency lowering buttons or the hand control (via T-cable).
HOT PLUGGING
Removing or adding any OpenBus™ cables are not allowed when the control box is on power via mains supply or battery! If necessary anyway
follow the below procedure:
1. Remove mains or battery and wait 5 sec.
2. Mount or dismount the required cables
If this procedure is NOT followed it may result in a damaged OpenBus™ driver circuit. The risk of a damaged circuit increases if the accessory has a
high start current (in rush current).
...to be continued
Warnings!
In order to avoid injury, the emergency-stop should be activated in (all) transport situations.
Functionality – JUMBO Care with display
Below you find information about what to read-out on the display version of JUMBO Care. Basically the functionality for the display version is the
same as the diode version, but more information can be read out on the display.