Page 68 of 108
6. Information on specific controls
The ACC (Attendant Control Compact) is fitted to advanced hospital and patient care beds for use where
patient positioning must be carefully controlled by medical staff.
1. ACC (MEDLINE
®
CARELINE
®
)
The OpenBus
TM
system makes it possible to use keypads as Attendant Controls or Handsets integrated in the
bed side rails. The keypads are named ACK (Attendant Control Keypads).
The protection class is dependent on customer design and testing.
2. ACK (MEDLINE
®
CARELINE
®
)
The ACL (Attendant Control Lock) box is a one turn button box fitted to hospital and care beds for use where
the patient positioning must be carefully controlled by the medical staff.
The ACL disconnects all functions on some handsets.
3. ACL (MEDLINE
®
CARELINE
®
)
Features and Options:
• There is no IP rating as standard, please contact LINAK A/S if a special IP rating is required
Usage:
• Compatibible with OpenBus™ control boxes e.g. CB20
• Ambient temperature: +5˚ to + 40˚C
• For approvals information see Lintra/Development/Certificates
For LINAK Standard ACK’s, the following is applicable:
• Colour: Grey RAL 7035
• Adhesive for the standard ACK is 3M 468MP
• For information re. suitable and unsuitable surfaces, please see 3M’s webpage
• Standard recommandation for curing time is 72 hours
• The customer is responsible for correct mounting on suitable surfaces
For datasheets and additional informations regarding the adhesives: www.3M.com
For Customised ACK’s, the following is applicable:
Dependant on the customer’s own design, installation, application and test it is possible to improve the IP protection class by the use of a special ACK
introducing an extra sealing ring to prevent against ingress
Mounting and curing time is dependant on customer’s application.
For datasheets and additional information regarding the adhesives: www.3M.com
Recommendations:
• The customer is responsible for the cable strain relief.
• Precautions against ESD (Electrostatic discharge) should be taken, as the PCBA is delivered separately.
• ESDS devices must under no circumstances, during transport, storage, handling, production or mounting in an application, be exposed to harmful ESD.
• The customer is responsible for correct mounting of the ACK on suitable surfaces.
• Keypads, PCBA and cables are supplied separately and have to be mounted by the customer.