Page 92 of 108
The Under Bed Light (UBL2) provides a powerful light with
a good distribution. The UBL2 is to be used for beds within
hospitals, nursing homes and in homecare.
The Under Bed Light makes it easier for patients and other
people in need of care to find their way at night in the dark
to prevent falling accidents and to make them feel safe.
9. Under Bed Light 2 (MEDLINE
®
CARELINE
®
)
Usage:
• Operating temperature: +5°C to +40°C
• Storage temperature: -10°C to +50°C
• Relative humidity: 10% to 80% at +30°C not condensing
• Atmospheric pressure: 700 to 1060hPa
• Approvals: EN62471, IEC60601-1:2005, 3
rd
edition (approvals are currently pending)
Recommendations & precautions:
• Screw holes in application are needed for mounting. Inform the customer to use M4 Ø12 screws with Ø12 washer, when mounting the UBL2.
Max. torque 2.5 Nm.
• Hot-plugging:
Removing or adding any OpenBus™ cables is not allowed when the CB is powered by mains supply!
If needed anyway, follow the below procedure:
1. Remove mains and wait 5 sec.
2. Mount or dismount the required cables
If this procedure is NOT followed, it may result in a damaged OpenBus™ driver circuit. The risk of a damaged circuit increases if the accessory
has a high start current (in rush current).
• There can be a risk of conflict with other OPENBUS™ accessories, like HB, ACP, etc. when using the OpenBus™ UBL2, it is therefore
recommended to make a system/bit overview.
• Always use locking mechanism and O-ring
• Sockets not used must be fitted with blind plugs to ensure IP Degree
• The UBL2 must be mounted on a plane surface and casing must not be subject to impact or any kind of stress.